Novel Method of Extraction for Radiocarbon Measurements of Atmospheric Carbon dioxide

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Original languageEnglish
Title of host publicationNovel Method of Extraction for Radiocarbon Measurements of Atmospheric Carbon dioxide
Pages1867-1877
Number of pages10
DatePublished - 28 Nov 2019

Abstract

In this paper, we present the first data from an alternative extraction method for atmospheric 14CO2 analysis, based on the direct trapping of whole air samples onto a molecular sieve zeolite (13X) trap, incorporated into a commercially available automated graphitization system. Results are presented for both inter-laboratory comparison samples and an in-house reference standard. The in-house reference was used to calculate the standard deviation of measurements (2.0‰). This newly developed method will facilitate faster sample processing and therefore lower cost per analysis, critical for scaling up such studies.