Quality assessment of anterior segment OCT images: development and validation of quality criteria

Radhika Pooja Patel, Harry Petrushkin, Katie Etherton, Katherine Terence, Andrew D Dick, Jugnoo S Rahi, Ameenat Lola Solebo

Research output: Contribution to journalArticle (Academic Journal)peer-review

1 Citation (Scopus)

Abstract

BACKGROUND: The utility of medical imaging is dependent on image quality. We aimed to develop and validate quality criteria for ocular anterior segment optical coherence tomography (AS-OCT) images.

METHODS: We undertook a cross-sectional study using AS-OCT images from patients aged 6-16. A novel three-level grading system (good, limited or poor) was developed based on the presence of image artefact (categorised as lid, eyelash, cropping, glare, or movement artefact). Three independent experts graded 2825 images, with agreement assessed using confusion matrices and intraclass correlation coefficients (ICC) for each parameter.

RESULTS: There was very good inter-grader IQA agreement assessing image quality with ICC 0.85 (95%CI: 0.84-0.87). The most commonly occurring artefact was eyelash artefact (1008/2825 images, 36%). Graders labelled 621/2825 (22%) images as good and 384 (14%) as poor. There was complete agreement at either end of the confusion matrix with no 'good' images labelled as 'poor' by other graders, and vice versa. Similarly, there was very good agreement when assessing presence of lash (0.96,0.94-0.98), movement (0.97,0.96-0.99), glare (0.82,0.80-0.84) and cropping (0.90,0.88-0.92).

CONCLUSIONS: The novel image quality assessment criteria (IQAC) described here have good interobserver agreement overall, and excellent agreement on the differentiation between 'good' and 'poor' quality images. The large proportion of images graded as 'limited' suggests the need for refine this classification, using the specific IQAC features, for which we also report high interobserver agreement. These findings support the future potential for wider clinical and community care implementation of AS-OCT for the diagnosis and monitoring of ocular disease.

Original languageEnglish
Article number103886
JournalPhotodiagnosis and photodynamic therapy
Volume45
Early online date20 Dec 2023
DOIs
Publication statusPublished - 1 Feb 2024

Bibliographical note

Copyright © 2023. Published by Elsevier B.V.

Fingerprint

Dive into the research topics of 'Quality assessment of anterior segment OCT images: development and validation of quality criteria'. Together they form a unique fingerprint.

Cite this