Skip to content

Secondary fluorescence in WDS: the role of spectrometer positioning

Research output: Contribution to journalArticle

Original languageEnglish
Number of pages8
JournalMicroscopy and Microanalysis
Early online date3 Dec 2018
DOIs
DateAccepted/In press - 16 Oct 2018
DateE-pub ahead of print (current) - 3 Dec 2018

Abstract

Secondary fluorescence, typically a minor error in routine electron probe microanalysis (EPMA), may not be negligible when performing high precision trace element analyses in multiphase samples. Other factors, notably wavelength dispersive spectrometer defocusing, may introduce analytical artefacts.
To explore these issues, we measured EPMA transects across two material couples chosen for their high fluorescence yield. We measured transects away from the fluorescent phase, and at various orientations with respect to the spectrometer focal line. Compared to calculations using both the Monte Carlo simulation code PENEPMA and the semi-analytical model FANAL, both codes estimate the magnitude of SF, but accurate correction requires knowledge of the position of the spectrometer with respect to the couple interface. Positioned over the fluorescent phase or otherwise here results in a factor of 1.2-1.8 of apparent change in SF yield.
SF and spectrometer defocusing may introduce systematic errors into trace element analyses, both may be adequately accounted for by modelling. Of the two, however, SF is the dominant error, resulting in 0.1 wt% Zn apparently present in Al at 100 m away from the Zn boundary in an Al/Zn couple. Of this, around 200 ppm Zn can be attributed to spectrometer defocusing.

    Research areas

  • secondary fluorescence, wavelength dispersive spectrometer, defocusing, electron microprobe, spectrometer positioning

Download statistics

No data available

Documents

Documents

  • Full-text PDF (accepted author manuscript)

    Rights statement: This is the author accepted manuscript (AAM). The final published version (version of record) is available online via Cambridge University Press at https://www.cambridge.org/core/journals/microscopy-and-microanalysis/article/secondary-fluorescence-in-wds-the-role-of-spectrometer-positioning/94F6F5D3992B37BFBB8B4116BB4605D3 . Please refer to any applicable terms of use of the publisher.

    Accepted author manuscript, 1.94 MB, PDF document

DOI

View research connections

Related faculties, schools or groups