Secondary fluorescence in WDS: the role of spectrometer positioning

Benjamin Buse, Jon Wade, Xaviet Llovet, Stuart Kearns

Research output: Contribution to journalArticle (Academic Journal)peer-review

7 Citations (Scopus)
367 Downloads (Pure)

Abstract

Secondary fluorescence, typically a minor error in routine electron probe microanalysis (EPMA), may not be negligible when performing high precision trace element analyses in multiphase samples. Other factors, notably wavelength dispersive spectrometer defocusing, may introduce analytical artefacts.
To explore these issues, we measured EPMA transects across two material couples chosen for their high fluorescence yield. We measured transects away from the fluorescent phase, and at various orientations with respect to the spectrometer focal line. Compared to calculations using both the Monte Carlo simulation code PENEPMA and the semi-analytical model FANAL, both codes estimate the magnitude of SF, but accurate correction requires knowledge of the position of the spectrometer with respect to the couple interface. Positioned over the fluorescent phase or otherwise here results in a factor of 1.2-1.8 of apparent change in SF yield.
SF and spectrometer defocusing may introduce systematic errors into trace element analyses, both may be adequately accounted for by modelling. Of the two, however, SF is the dominant error, resulting in 0.1 wt% Zn apparently present in Al at 100 m away from the Zn boundary in an Al/Zn couple. Of this, around 200 ppm Zn can be attributed to spectrometer defocusing.
Original languageEnglish
Number of pages8
JournalMicroscopy and Microanalysis
Early online date3 Dec 2018
DOIs
Publication statusE-pub ahead of print - 3 Dec 2018

Keywords

  • secondary fluorescence
  • wavelength dispersive spectrometer
  • defocusing
  • electron microprobe
  • spectrometer positioning