Surgical experience with an implanted artificial cervical joint

Brian H Cummins, Steven S Gill, James T. Robertson

Research output: Contribution to journalArticle (Academic Journal)peer-review

197 Citations (Scopus)

Abstract

Object. To assess the effectiveness of Cummins' artificial cervical joint, the authors reviewed the cases of 20 patients in whom the joint had been placed.

Methods. A review of patients' medical records and reexamination of 18 patients were performed. The review of the surgical experience with the implantation of movable stainless-steel joints in 20 patients treated for cervical myelopathy (16 patients), cervical radiculopathy (three patients), and severe pain (one patient) indicated that the procedure is safe and well tolerated and does preserve cervical joint motion in most patients over an extended period of observation. To date, adjacent segmental symptomatic degenerative changes leading to further surgical treatment have been avoided. The joint has been placed in patients with advanced congenital and acquired cervical fusion and has been demonstrated to be stable, mobile, and biomechanically and biochemically compatible; it has shown no subsidence into adjacent bone. Wear debris has not occurred.

Conclusions. The use of stainless steel in the cervical spine appears to be suitable for this joint replacement design.
Original languageEnglish
Pages (from-to)943-948
Number of pages6
JournalJournal of Neurosurgery
Volume88
Issue number6
Publication statusPublished - Jun 1998

Fingerprint

Dive into the research topics of 'Surgical experience with an implanted artificial cervical joint'. Together they form a unique fingerprint.

Cite this