Abstract
BACKGROUND: In the prospective, multicentre, randomised TARGET All Comers study, percutaneous coronary intervention (PCI) with the FIREHAWK biodegradable-polymer sirolimus-eluting stent (BP-SES) was non-inferior to the durable-polymer everolimus-eluting stent (DP-EES) for the primary endpoint of target lesion failure (TLF) at 12 months.
AIMS: We aimed to report the final study outcomes at 5 years.
METHODS: Patients referred for PCI were randomised to receive either a BP-SES or DP-EES in a 1:1 ratio in 10 European countries. Randomisation was stratified by centre and ST-elevation myocardial infarction (STEMI) presentation, and clinical follow-up extended to 5 years. The primary endpoint was TLF (composite of cardiac death, target vessel myocardial infarction [MI], or ischaemia-driven target lesion revascularisation). Secondary endpoints included patient-oriented composite events (POCE; composite of all-cause death, all MI, or any revascularisation and its components).
RESULTS: From December 2015 to October 2016, 1,653 patients were randomly assigned to the BP-SES or DP-EES groups, of which 93.8% completed 5-year clinical follow-up or were deceased. At 5 years, TLF occurred in 17.1% of the BP-SES group and in 16.3% of the DP-EES group (p=0.68). POCE occurred in 34.0% of the BP-SES group and 32.7% of the DP-EES group (p=0.58). Revascularisation was the most common POCE, occurring in 19.3% of patients receiving BP-SES and 19.2% receiving DP-EES, of which less than one-third was ischaemia-driven target lesion-related. In the landmark analysis, there were no differences in the rates of TLF and POCE between groups from 1 to 5 years, and these results were consistent across all subgroups.
CONCLUSIONS: In an all-comers population requiring stent implantation for myocardial ischaemia, the BP-SES was non-inferior to the DP-EES for the primary endpoint of TLF at 12 months, and results were sustained at 5 years, confirming the long-term safety and efficacy of the FIREHAWK BP-SES.
| Original language | English |
|---|---|
| Pages (from-to) | e844-e855 |
| Journal | EuroIntervention |
| Volume | 19 |
| Issue number | 10 |
| DOIs | |
| Publication status | Published - 4 Dec 2023 |
Bibliographical note
Funding Information:Funding for our study was provided by Shanghai MicroPort Medical (Group) Co., Ltd.
Publisher Copyright:
© Europa Digital & Publishing 2023. All rights reserved.
Keywords
- Humans
- Sirolimus
- Drug-Eluting Stents/adverse effects
- Percutaneous Coronary Intervention/adverse effects
- Prospective Studies
- Treatment Outcome
- Absorbable Implants
- Everolimus
- Myocardial Infarction/etiology
- Polymers