Abstract
In the process of a liquid chromatography–tandem mass spectrometry (LC–MS/MS) assay development for two antimicrobial drugs using their stable isotope-labeled (SIL) analogues as internal standards to correct for matrix–ion suppression effects, an unusually large scatter of data was observed. A systematic observation revealed that the analytes and their SIL analogues (internal standards) were not completely coeluted; therefore, the analytes were not experiencing matrix effects similarly. When a column with lower resolution ability was used to achieve complete overlapping of the analyte and internal standard peaks, the scatter of LC–MS/MS data was minimized, indicating that the maximum correction of matrix effects by internal standards occurs when they are completely coeluted with the analytes. This study highlights the importance of ensuring complete overlapping of analyte and internal standard peaks in eliminating matrix effects when using SIL analogues as internal standards in LC–MS.
Original language | English |
---|---|
Pages (from-to) | 335-338 |
Number of pages | 4 |
Journal | LC-GC North America |
Volume | 39 |
Issue number | 7 |
Publication status | Published - 1 Jul 2021 |
Bibliographical note
Funding Information:This study was funded by an International Association for Dental Research (IADR)-GlaxoSmithKline Innovation in Oral Care Award 2016 awarded to Dr. H.M.H.N. Bandara.
Funding Information:
This study was funded by an Interna tional Association for Dental Research (IADR)-GlaxoSmithKline Innovation in Oral Care Award 2016 awarded to Dr. H.M.H.N. Bandara.
Publisher Copyright:
© 2021, UBM Medica Healthcare Publications. All rights reserved.